Mitral valve-in-valve: crossing a calcified interatrial septum

Transcatheter mitral valve-in-valve replacement (TMVR) in patients p...

The challenge of mitral Valve-in-Valve in hostile territory

Transcatheter mitral valve replacement (TMVR) "valve-in-valve" in patients with a history of complex cardiac surgery represents a major technical challenge, particularly when the septal anatomy is compromised. This study reports the case of a 55-year-old patient, initially operated on for an incomplete atrioventricular canal with implantation of a Carpentier PERIMOUNT 27 bioprosthesis, presenting with severe prosthetic degeneration associated with a heavily calcified interatrial septum.

The objective of this intervention was to perform a valve-in-valve implantation despite the impossibility of using a conventional transseptal puncture due to pathological septal rigidity. The clinical hypothesis tested here relied on the feasibility of a septal crossing strategy assisted by a Versacross system, followed by the creation of an arteriovenous loop to stabilize the placement of a SAPIEN 3 Ultra 26 prosthesis. This report documents not only the resolution of this extreme anatomical constraint but also confirms the functional durability of the valve at one-year follow-up, offering a concrete alternative to iterative surgical options in high-risk patients.

Methodology and clinical management

This case report describes a mitral valve-in-valve procedure in a 55-year-old patient with a history of surgical repair of an incomplete atrioventricular canal, with implantation of a Carpentier PERIMOUNT 27 bioprosthesis (Edwards, USA).

The intervention protocol had to overcome a major difficulty: the impossibility of performing a conventional transseptal puncture due to a severely calcified interatrial patch. The procedure followed these steps:

  • Access and crossing: Use of the Versacross system (Boston Scientific) to successfully perform the transseptal puncture through the calcified patch.
  • Navigation: Placement of a deflectable catheter to advance a high-performance guidewire, captured in the left ventricle to establish a veno-arterial loop.
  • Preparation: Septostomy performed using an Atlas Gold balloon catheter of 16 mm × 45 mm (BD, USA).
  • Implantation: Introduction of a 26-Fr DRYSEAL sheath (Gore, USA) into the left atrium, followed by the implantation of a SAPIEN 3 Ultra 26 bioprosthesis (Edwards, USA) with nominal inflation +2 cm³.

The evaluation of technical success was based on intraoperative imaging and clinical follow-up including a 1-year assessment, confirming the long-term stability of valve function.

Management of a degenerated mitral bioprosthesis on a calcified septum: a complex approach

The reported case concerns a 55-year-old patient, carrier of a mitral bioprosthesis (Carpentier PERIMOUNT 27) following surgical correction of an incomplete atrioventricular canal. Faced with progressive prosthetic degeneration, a transcatheter mitral valve-in-valve procedure was decided upon.

The major difficulty lay in the severe calcification of the interatrial patch, making conventional transseptal puncture impossible. The team had to deploy a sophisticated interventional strategy:

  • Access: Use of the Versacross system to cross the calcified patch.
  • Technique: Establishment of a veno-arterial loop using a deflectable catheter and a high-support guidewire, captured in the left ventricle.
  • Preparation: Septostomy performed with an Atlas Gold balloon (16 mm × 45 mm).
  • Implantation: Placement of a 26-Fr DRYSEAL sheath to deliver a SAPIEN 3 Ultra 26 valve, deployed with nominal inflation +2 cm³.

The immediate and medium-term results demonstrate the feasibility of this approach despite a hostile anatomical environment:

ParameterResult
Implanted deviceSAPIEN 3 Ultra 26
Prosthetic expansionOptimal, without conflict with the left ventricular outflow tract
Clinical follow-upNormal valve function at 1 year

Post-procedural imaging confirmed excellent expansion of the bioprosthesis, with slight protrusion into the left ventricle without outflow tract obstruction. The patient maintained satisfactory prosthetic function during the 12-month evaluation.

Management of the calcified interatrial septum: a complex transcatheter approach

The clinical case reported by Fernández González et al. illustrates a rescue strategy for a 55-year-old patient with a mitral bioprosthesis (Carpentier PERIMOUNT 27) presenting with severe prosthetic degeneration. The major complexity lay in the impossibility of conventional transseptal access due to a heavily calcified interatrial septum, sequelae of a previous surgical correction of an incomplete atrioventricular canal.

The success of the intervention relied on the use of the Versacross system to cross the calcified area, followed by the creation of an arteriovenous loop. Performing a septostomy using an Atlas Gold balloon (16 mm × 45 mm) allowed for the secure passage of a 26-Fr DRYSEAL sheath. Successful implantation of a SAPIEN 3 Ultra valve (26 mm) restored optimal prosthetic function, confirmed by one-year follow-up.

Clinical limitations and perspectives

Although this report demonstrates the technical feasibility of mitral valve-in-valve in a hostile environment, it is a single observation (N=1). This approach highlights the need for rigorous planning in cases of complex surgical history, where massive calcifications transform a standard procedure into a major technical challenge. Mastery of septal crossing tools is decisive here to avoid iatrogenic complications, as observed during the initial transseptal puncture attempt.

Case summary: Mitral valve-in-valve on calcified septum

A 55-year-old patient with a mitral bioprosthesis (Carpentier PERIMOUNT 27) following atrioventricular canal correction underwent a percutaneous valve-in-valve intervention. The procedure required a transseptal puncture assisted by the Versacross system through a severely calcified patch, followed by a septostomy with an Atlas Gold balloon (16 mm × 45 mm) to allow the introduction of a 26-Fr sheath and the implantation of a SAPIEN 3 Ultra prosthesis (26 mm). Technical success was confirmed by normal prosthetic function at 12 months.

Concretely, for the practitioner:

  • Management of difficult access: The use of the Versacross system and a veno-arterial loop (by capturing a guidewire in the LV) allows crossing highly calcified interatrial patches where conventional transseptal puncture fails.
  • Anatomical complexity: Prior dilation with a large-caliber balloon is crucial to secure the delivery of large-diameter systems (26-Fr) in surgically altered areas.
  • Long-term results: The transseptal valve-in-valve technique constitutes a viable and durable therapeutic option, even in patients with complex cardiac history, provided there is meticulous preparation of the septum.
The therapeutic challenge reported by Fernández González et al. (2026) illustrates a complex situation of mitral bioprosthesis (Carpentier PERIMOUNT 27) degeneration in a 55-year-old patient, initially operated on for a complete atrioventricular canal. The particularity here lies in the management of a severely calcified interatrial septum which precluded a conventional transseptal puncture. To overcome this obstacle, the team used the Versacross system to cross the calcified patch, supplemented by the establishment of a veno-arterial loop allowing the placement of a 26-Fr DRYSEAL introducer in the left atrium. A SAPIEN 3 Ultra 26 prosthesis was then successfully deployed. The 1-year check-up confirms the durability of this approach, with preserved prosthetic function, despite an anatomy initially described as hostile. Concretely, for the practitioner: - Severe calcification of an interatrial patch is not an absolute contraindication to mitral Valve-in-Valve, provided that dedicated puncture equipment (Versacross system) is available. - The establishment of a veno-arterial loop is a robust strategy to secure the advancement of large-caliber devices (26-Fr) in complex anatomical conditions. - Long-term follow-up remains essential to monitor the stability of the prosthesis implanted in this scarred and calcified environment.

Source

  • Original title: Mitral valve-in-valve with severe atrial septum calcification
  • Authors: Luís Fernández González, Koldobika García San Román, Juan Carlos Astorga Burgo, Aída Acín Labarta, A. J. Reynolds, Roberto Blanco Mata
  • Publication: REC interventional cardiology (English Edition) - 2025-08-04
  • DOI: https://doi.org/10.24875/recice.m25000530

Information intended for healthcare professionals. This content may contain errors or truncated summaries. We recommend always verifying with the original source article. Delynov disclaims any liability regarding the use of this information. This document is not intended for patients or the general public.

M-TEER: estimated plasma volume predicts 5-year mortality
Fluid overload and systemic congestion are major determinants of morbidity
CardiacDistributor in France for

SERAMON® CHORDAE LOOP (exclusive), SERAMON® robotic and PTFE felt pledgets, manufactured in Naila (Germany). A team based in Épinal, France.

Implantology and periodontology ›

Legal notice | Privacy

The medical devices presented on this site, SERAMON® CHORDAE LOOP, SERAMON® robotics and PTFE felt pledgets, are manufactured by SERAG-WIESSNER GmbH & Co. KG (Naila, Germany) and distributed in France by Delynov. For healthcare professionals only; read the instructions for use carefully before use: indications, contraindications, adverse effects and precautions. SERAMON® CHORDAE LOOP is a class III implantable medical device, sterile, for single use; additional knots may be required on tissue under tension. Any serious incident must be reported to the manufacturer and the ANSM. Animations and instruments shown: generic 3D models designed by Delynov, for illustration purposes only. Electronic instructions for use: ifu.serag-wiessner.de. Ref. DLC-SITE, version 12 of 06/10/2026.
© 2026 Delynov SASFor healthcare professionals only